===== Status ===== | ===== Status ===== | ||
- | |||
- | == 02 September 2015 == | ||
- | Chiller water turned black due to the break down of glycol in the cooling water (causing the cooling water to become corrosive). | ||
== 12 August 2014 == | == 12 August 2014 == | ||
- | Leak detected between bell jar and base plate. | + | **Inactive** - Leak detected between bell jar and base plate. |
Line 30: | Line 27: | ||
===== Maintenance Log===== | ===== Maintenance Log===== | ||
- | 06/19/2014 - Diffusion pump cleaned, and oil changed. - JH | + | == 09/02/2015 - JH == |
+ | |||
+ | - Chiller water turned black due to the break down of glycol in the cooling water (causing the cooling water to become corrosive). | ||
+ | |||
+ | - Update to the waterlines for cooling the evaporation terminals and crystal monitors is complete. | ||
+ | |||
+ | - Water leak found for line going into diffusion pump, possibly caused by the corrosive water in water the lines. | ||
+ | |||
+ | == 06/ | ||
+ | - Diffusion pump cleaned, and oil changed. | ||
+ | |||
+ | == 06/ | ||
+ | |||
+ | - Roughing pump oil changed. Old oil in good condition. | ||
+ | |||
+ | == 07/13/2012 - RW == | ||
+ | |||
+ | - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. | ||
- | 06/19/2013 - Roughing pump oil changed. | + | == 07/03/2012 - RW == |
+ | |||
+ | - Roughing pump oil changed. | ||
- | 07/13/2012 - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. | + | == 8/09/2011 - RW == |
- | 07/ | + | - Roughing pump oil changed. |
- | 8/09/2011 - Roughing pump oil changed. Old oil was brown and had significant amounts of debris floating in it. Added a permanent maintenance log to the blue binder holding all of the user manuals for the system. | + | == 6/17/2011 - RW == |
- | 6/ | + | - Inside of bell jar and all surfaces cleaned. Shutter operation fixed. Ion gauge degaussed; now reads more realistic pressure. |
====== Schrader Thermal Evaporator Wngr 489 ====== | ====== Schrader Thermal Evaporator Wngr 489 ====== |