Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
thermal:start [2013/06/21 13:49] – janet | thermal:start [2020/03/06 09:04] (current) – external edit 127.0.0.1 | ||
---|---|---|---|
Line 3: | Line 3: | ||
===== Status ===== | ===== Status ===== | ||
- | = 05 July 2012 = | + | |
+ | == 12 August 2014 == | ||
+ | **Inactive** - Leak detected between bell jar and base plate. | ||
+ | |||
+ | |||
+ | == 05 July 2012 == | ||
Two of the three crystal monitors are currently operational. | Two of the three crystal monitors are currently operational. | ||
Shutter is not operational. | Shutter is not operational. | ||
Line 22: | Line 27: | ||
===== Maintenance Log===== | ===== Maintenance Log===== | ||
- | 07/13/2012 - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. - RW | + | == 09/02/2015 - JH == |
+ | |||
+ | - Chiller water turned black due to the break down of glycol in the cooling water (causing the cooling water to become corrosive). | ||
+ | |||
+ | - Update to the waterlines for cooling the evaporation terminals and crystal monitors is complete. | ||
+ | |||
+ | - Water leak found for line going into diffusion pump, possibly caused by the corrosive water in water the lines. | ||
+ | |||
+ | == 06/19/2014 - JH == | ||
+ | - Diffusion pump cleaned, and oil changed. | ||
+ | |||
+ | == 06/19/2013 - RW == | ||
+ | |||
+ | - Roughing pump oil changed. Old oil in good condition. | ||
+ | |||
+ | == 07/ | ||
+ | |||
+ | - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. | ||
+ | |||
+ | == 07/03/2012 - RW == | ||
+ | |||
+ | - Roughing pump oil changed. Good color. Some debris. | ||
+ | |||
+ | == 8/ | ||
- | 07/ | + | - Roughing pump oil changed. |
- | 8/09/2011 - Roughing pump oil changed. Old oil was brown and had significant amounts of debris floating in it. Added a permanent maintenance log to the blue binder holding all of the user manuals for the system. | + | == 6/17/2011 - RW == |
- | 6/ | + | - Inside of bell jar and all surfaces cleaned. Shutter operation fixed. Ion gauge degaussed; now reads more realistic pressure. |
====== Schrader Thermal Evaporator Wngr 489 ====== | ====== Schrader Thermal Evaporator Wngr 489 ====== | ||
Stainless steel box evaporator commissioned 1991. Conflat-sealed except for o-ring-sealed front door. Base pressure few x 10^-8 torr. Turbo pumped. | Stainless steel box evaporator commissioned 1991. Conflat-sealed except for o-ring-sealed front door. Base pressure few x 10^-8 torr. Turbo pumped. | ||
- | ===== Current | + | ===== Status ===== |
+ | == June 2013 == | ||
Upgraded over 2012/13 academic year by Casey Hines. Casey' | Upgraded over 2012/13 academic year by Casey Hines. Casey' | ||
- | ==== Status - June 2011 ==== | + | == June 2011 == |
Inactive. | Inactive. | ||
===== Maintenance Log===== | ===== Maintenance Log===== |