Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
thermal:start [2013/06/21 13:42] – janet | thermal:start [2020/03/06 09:04] (current) – external edit 127.0.0.1 | ||
---|---|---|---|
Line 1: | Line 1: | ||
====== Veeco Thermal Evaporator Wngr 489 ====== | ====== Veeco Thermal Evaporator Wngr 489 ====== | ||
- | Glass bell jar pumped by a diffusion pump which is backed by a mechanical pump. Commissioned about 1965? O-ring sealed. Base pressure about 1 x 10^-6 torr. Ion gauge and convectron gauges. | + | Glass bell jar pumped by a diffusion pump backed by a mechanical pump. Commissioned about 1965? O-ring sealed. Base pressure about 1 x 10^-6 torr. Ion gauge and convectron gauges. |
- | ==== Current | + | ===== Status |
+ | == 12 August 2014 == | ||
+ | **Inactive** - Leak detected between bell jar and base plate. | ||
+ | |||
+ | |||
+ | == 05 July 2012 == | ||
Two of the three crystal monitors are currently operational. | Two of the three crystal monitors are currently operational. | ||
Shutter is not operational. | Shutter is not operational. | ||
Line 22: | Line 27: | ||
===== Maintenance Log===== | ===== Maintenance Log===== | ||
- | 07/13/2012 - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. - RW | + | == 09/02/2015 - JH == |
- | 07/03/2012 - Roughing pump oil changed. Good color. Some debris. - RW | + | - Chiller water turned black due to the break down of glycol in the cooling water (causing the cooling water to become corrosive). |
- | 8/ | + | - Update |
- | 6/ | + | - Water leak found for line going into diffusion pump, possibly caused by the corrosive water in water the lines. |
- | ====== Schrader Thermal Evaporator Wngr 489 ====== | + | == 06/19/2014 - JH == |
- | Stainless steel box evaporator commissioned 1991. Conflat-sealed except for o-ring-sealed front door. Base pressure few x 10^-8 torr. Turbo pumped. | + | - Diffusion |
- | Upgraded over 2012/13 academic year by Casey Hines. Casey' | + | == 06/19/2013 - RW == |
+ | |||
+ | - Roughing pump oil changed. Old oil in good condition. | ||
+ | |||
+ | == 07/13/2012 - RW == | ||
+ | |||
+ | - Diffusion pump oil changed. Refilled with 200 cc Dow 704. Old oil had yellowish color. No visible debris. Some tarring on the DP walls. Fixed minor leaks in deposition rate detector water lines. Chamber now holds at 0.1 Torr for 3+ days. | ||
+ | |||
+ | == 07/03/2012 - RW == | ||
+ | |||
+ | - Roughing pump oil changed. Good color. Some debris. | ||
+ | |||
+ | == 8/09/2011 - RW == | ||
+ | |||
+ | - Roughing pump oil changed. Old oil was brown and had significant amounts of debris floating in it. Added a permanent maintenance log to the blue binder holding all of the user manuals for the system. | ||
+ | |||
+ | == 6/17/2011 - RW == | ||
+ | |||
+ | - Inside of bell jar and all surfaces cleaned. Shutter operation fixed. Ion gauge degaussed; now reads more realistic pressure. | ||
+ | |||
+ | ====== Schrader Thermal Evaporator Wngr 489 ====== | ||
+ | Stainless steel box evaporator commissioned 1991. Conflat-sealed except for o-ring-sealed front door. Base pressure few x 10^-8 torr. Turbo pumped. | ||
+ | ===== Status ===== | ||
+ | == June 2013 == | ||
+ | Upgraded over 2012/13 academic year by Casey Hines. Casey' | ||
- | ===== Current Status (April | + | == June 2011 == |
Inactive. | Inactive. | ||
===== Maintenance Log===== | ===== Maintenance Log===== |