seebeck:troubleshooting [2018/04/16 14:15] – OFLO error troubleshooting suggestions for Keithley 195A aaron | seebeck:troubleshooting [2020/03/06 09:04] (current) – external edit 127.0.0.1 | ||
---|---|---|---|
Line 14: | Line 14: | ||
1. Double check the connection to the computer at the back of the Keithley. | 1. Double check the connection to the computer at the back of the Keithley. | ||
+ | |||
2. Turn the device off. After 20 minutes or more turn the device back on and wait the 1 hour warm up time. | 2. Turn the device off. After 20 minutes or more turn the device back on and wait the 1 hour warm up time. | ||
+ | |||
3. (This one might just be luck but it's worked twice for me now after following the above steps) Ensure that the Keithley connected to the Dim 5120 PDMM is turned on. It's not connected to the Keithley 195A at all but I've noticed it works. I would guess that Lab View expects that DMM to be there and there are issues if it's not turned on. This is just a hypothesis though and needs further testing. | 3. (This one might just be luck but it's worked twice for me now after following the above steps) Ensure that the Keithley connected to the Dim 5120 PDMM is turned on. It's not connected to the Keithley 195A at all but I've noticed it works. I would guess that Lab View expects that DMM to be there and there are issues if it's not turned on. This is just a hypothesis though and needs further testing. | ||