 6) Make sure to record the position and intensity (# counts) of the reference Si calibration peak every time calibration file on the Raman computer desktop. Also your name and the date. This helps to track the system folder. 65,000 to 70,000 counts is high; more usually 45,000. Should be at 520 cm^-1, can vary to 518-523 cm^-1 (temperature?). 6) Make sure to record the position and intensity (# counts) of the reference Si calibration peak every time calibration file on the Raman computer desktop. Also your name and the date. This helps to track the system folder. 65,000 to 70,000 counts is high; more usually 45,000. Should be at 520 cm^-1, can vary to 518-523 cm^-1 (temperature?).
 +
 +7) Usual Raman mode is 3s accumulation and 3 times at each energy. Can increase time and frequency.
  