pld:start [2016/04/14 10:40] jamespld:start [2016/04/14 11:21] james
Line 72: Line 72:
 There are many windows on the thermionics chamber.  However, there are two UV grade laser windows.  Each of these is mounted in a ConFlat flange.  Each of the laser windows is made of [[http://marketplace.idexop.com/store/SupportDocuments/MaterialProperties.pdf|fused silica]], and measuring with the Coherent power meter, have a loss of approximately 10% at 248nm. There are many windows on the thermionics chamber.  However, there are two UV grade laser windows.  Each of these is mounted in a ConFlat flange.  Each of the laser windows is made of [[http://marketplace.idexop.com/store/SupportDocuments/MaterialProperties.pdf|fused silica]], and measuring with the Coherent power meter, have a loss of approximately 10% at 248nm.
  
-===== Procedure ===== +===== Standard Operating Procedure ===== 
-{{:pld:pld_procedure.tex.pdf|Thermionics Chamber}} - Josh Russell+{{:pld:pld_procedure.tex.pdf|Thermionics Chamber}}
  
-{{:pld:neocera.pld.instructions.pdf|Neocera Chamber}}+{{:pld:neocera_chamber_sop.pdf|Neocera Chamber}} 
 + 
 +{{:pld:laser_sop-tate_lab.doc|Laser Operation}}
  
 {{:pld:gas_change_20150416.pdf|Laser Gas Change}} {{:pld:gas_change_20150416.pdf|Laser Gas Change}}
  
 {{:pld:PLD_viewing_window_cleaning_20150422.pdf|Chamber Window Cleaning}} {{:pld:PLD_viewing_window_cleaning_20150422.pdf|Chamber Window Cleaning}}
- 
- 
-FIXME 
-Chamber cleaning procedures 
  
 ===== Routine Maintenance ===== ===== Routine Maintenance =====
Line 104: Line 102:
 ==Changing Gas== ==Changing Gas==
 Process Tube {{:pld:os01000-001s_layout1_1_.pdf|gas cabinet layout}} Process Tube {{:pld:os01000-001s_layout1_1_.pdf|gas cabinet layout}}
-The premixed gas we use is: 0.1% F2, 1.7% He, 3.93% Kr, balance Ne (i.e. rest is Neon), and the SDS is {{:pld:dbc2-12_chamber_darwing.pdf|KrF premix SDS}}+The premixed gas we use is: 0.1% F2, 1.7% He, 3.93% Kr, balance Ne (i.e. rest is Neon), and the SDS is {{:pld:krf_premix_spectra.pdf|KrF premix SDS}}
  
-Usually purchased from Airgas, contact can be found here [[:Purchasing|Purchasing]]+Usually purchased from Airgas, contact can be found by following the [[:Purchasing|Purchasing]] link
  
  

QR Code